OMPLIANCE WITH ELECTROMAGNETIC EMISSION STANDARDS
The following paragraphs specify compliance with the electromagnetic emission standards:
▪
Guidelines and manufacturer declaration - Electromagnetic emissions
▪
Guidelines and manufacturer declaration - Electromagnetic immunity
▪
Recommended distances between portable and mobile radiofrequency communication equipment and the
AED
14.1
G
UIDELINES AND MANUFACTURER DECLARATION
-
E
LECTROMAGNETIC EMISSIONS
The
Saver ONE P
was designed to be used in electromagnetic environments with features listed below. The customer
or the user of the
Saver ONE P
must ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - Guidelines
RF emissions
CISPR 11
Group 1
The
AED
uses RF energy only for its internal operation. Its
RF emissions are, therefore, very low and it is improbable
that they may interfere with electronic devices nearby.
RF emissions
CISPR 11
Class B
The
AED
can be used in any building, including residential
buildings and buildings directly connected to the public
low-voltage electricity network that supplies residential
buildings.
Harmonic Emissions
IEC/EN 61000-3-2
Not applicable
Voltage
fluctuations/flickers
IEC/EN 61000-3-3
Not applicable
14.2
G
UIDELINES AND MANUFACTURER DECLARATION
-
E
LECTROMAGNETIC IMMUNITY
The
Saver ONE P
was designed to be used in electromagnetic environments with features listed below. The customer
or the user of the
Saver ONE P
must ensure that it is used in such an environment.
Immunity test
Test level
IEC/EN 60601-1
Conformity
Level
Electromagnetic environment
Guidelines
Electrostatic
discharge
(ESD)
IEC/EN 61000-4-2
±6 kV contact
±6 kV contact
The floors must be made of wood, cement or
ceramic bricks. If the floors are covered by
synthetic materials, the relative humidity must
be at least 30%.
±8 kV air
±8 kV air
Fast transients/burst
IEC/EN 61000-4-4
±2 kV for electricity
networks
Not applicable
±1 kV for I/O networks
±1 kV for I/O lines
IEC/EN 61000-4-4
5% U
T
(95% dip in U
T
)
for 0.5 cycles
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in
U
T
) for 5 seconds
Not applicable