Code
Displayed message
Description
Priority
Alarm signalling delay
(max)
1
Cardiac Low frequency
Detection of a slow sinus rhythm
HIGH
< 10 sec *
2
Asystole
Detection of an absent sinus rhythm
HIGH
< 10 sec *
3
Fibrillation Detected
Detection of ventricular fibrillation.
In this case the patient should undergo a
defibrillation shock
HIGH
< 5 sec
4
Tachycardia Detected
Detection of fast ventricular tachycardia. In
this case the patient should undergo a
defibrillation shock
HIGH
< 5 sec
5
Cardiac High frequency
Detection of an accelerated sinus rhythm. In
this case the patient should NOT undergo a
defibrillation shock
HIGH
< 5 sec
* see warning section for the use of the monitoring mode
The technical alarms detected are:
Code
Displayed message
Description
Priority
Alarm signalling delay
(max)
1
Patient lost
Absence of patient detection
HIGH
< 5 sec
2
ECG saturation
Non-operational device condition due to
saturation of the ECG amplifier stage
HIGH
< 5 sec