69
13.13
D
ISPLAY
Category
Rated specifications
Type
Colour TFT
with LED backlight
Visible area
5.7” (diagonal)
- 112 x 80 mm
Resolution
640 x 480 pixels
ECG tracing displayed
1 (Lead II)
ECG tracing speed
25 mm/sec (default)
Information shown on
Display
•Patient’s heart rate (bpm)
•Patient thoracic impedance (Ω)
•VF or VT detected (numeric value)
•Shocks delivered (numeric value)
• Textual prompts on operations to be performed (text)
• Graphical images on operations to be performed (graphic icons)
• Active technical and
physiological alarms (graphic icons)
•Operational mode
• Set energy level (J)
• Charge duration (graphic incremental bar)
• Treatment duration (hh/mm/sec)
• Battery level (incremental bar graph)
• Local date and time (dd/month/year
- hh/mm/sec)
13.14
N
ON
-
RECHARGEABLE BATTERY
Category
Rated specifications
REF (Model)
SAV-C0903
SAV-C0904
Type
Li-
SOCl
2
(Lithium-thionyl
chloride)
disposable, non-rechargeable
Li-MnO
2
(Lithium ions and Manganese
dioxide) disposable, non-rechargeable
Voltage - Capacity
25.2 VDC
–
3500 mAh
24 VDC
–
3000 mAh
Performance
*
Version 200J
250 complete rescue cycles (shocks at 200J. and CPR)
Version 360J
160 complete rescue cycles (shocks at 360J. and CPR)
Monitoring
ECG monitoring duration 24 hours continuously
Duration in Stand-by mode
(battery installed)
*
4 years if installed in the AED, assuming an activation test, daily self-tests without turning on
the AED
*New and fully charged battery stored at a constant temperature of 20°C and relative humidity of 45% without condensation
13.15
R
ECHARGEABLE BATTERY
Category
Rated specifications
REF (Model)
SAV-C0011
Type
Li ion (lithium ions) Rechargeable
Voltage - Capacity
21.6 VDC
–
2100 mAh
Performance
*
Version 200J
200 continuous shocks with new fully charged battery
Version 360J
110 continuous shocks with new fully charged battery
Monitoring
ECG monitoring duration 14 hours continuously
Charge time
*
≤
2.5 hours with new batteries and charging station SAV-C0012
Shelf Life
*
2 years or 300 charge/shock cycles (
whichever occurs first
)
*New and fully charged battery stored at a constant temperature of 20°C and relative humidity of 45% without condensation