5.5
2-
POLE
ECG
CABLE
SAV-C0017
The SAV-C0017 ECG cable is fitted with two clip terminals for disposable pre-gelled electrodes (
optional
).
The ECG cable can read the patient's lead II to show the ECG trace on the
Saver One P display.
The SAV-C0017 ECG
cable can only be used if the
Saver One P
is set in "ECG MONITORING" operating mode (see Chapter 10).
The ECG SAV-C0017 cable is classified as type CF
5.5.1
Placement of the electrodes
The SAV-C0017 ECG cable electrodes must be placed as shown in figure (19):
Electrode R
: near the right shoulder, directly below the clavicle.
Electrode F
: on the left side of the hypogastrium.
International coding
(European IEC/EN)
Code (IEC/EN)
Colour
(IEC/EN)
R
RED
F
GREEN