The infrared interface emits optically invisible radiation. The emission diode complies with IEC/EN 60825-1 Class 1 "Eye
Save".
➢
Do not use Paediatric Defibrillation PADs (SAV-C0016) on adult patients (older than 8 years and weighing more than
25kg). When using Paediatric Defibrillation PADs the
Saver One P
(Ref:
SVP-B0006, SVP-B0007
) automatically switches
to paediatric mode, reducing the maximum deliverable energy to 50J.
➢
Do not use SAV-C0599 Universal Face to Face Defibrillation PADs in paediatric mode on adult patients (older than 8
years and weighing over 25kg). By setting the paediatric mode in the menu, the
Saver One P
(Ref.
SVP-B0006-U, SVP-
B0006-U-Q, SVP-B0007-U, SVP-B0007-U-Q
) switches to paediatric mode, reducing the maximum deliverable energy to
50J.
➢
Place the patient cables so as to reduce the possibility of entangling or strangling the patient.
➢
In a domestic environment, keep the defibrillator away from the reach of children and pets.
➢
Do not apply the defibrillation electrodes directly on an implanted pacemaker to avoid any reading errors by the device
and to avoid damage to the pacemaker through the defibrillation pulse.
➢
Disconnect the patient from equipment that is sensitive to high voltage pulses, or equipment that is not defibrillator-
proof, before delivering the shock.
WARNING
➢
Do not allow the defibrillation electrodes to touch or to come into contact with ECG electrodes, pads, transdermal
plasters, etc. Otherwise, the formation of electric arcs and burns to the patient could be caused during defibrillation;
the current may even be dispersed.
➢
Place the defibrillation pads as indicated in this user manual and on the packaging.
➢
Do not use the defibrillation PADs if the gel has detached from the support or if it appears torn, detached or dry
➢
If damage has been observed, do not operate the
Saver One P
under any circumstances.
➢
Before using the device, remove metal objects from the patient's body (including necklaces or bracelets, etc.)
➢
Do not use defibrillation pads other than those supplied by the manufacturer. Otherwise, the defibrillator may perform
false interpretations.
➢
Do not use the defibrillation pads if they are damaged, even partly.
➢
Do not use defibrillation PADs if the expiry date has been exceeded.
➢
When applying the ECG SAV-C0017 cable make sure it is not in contact with any conductive element. Ensure that all
ECG electrodes are properly secured to the patient.
➢
Do not touch the patient or PADs during heart rhythm analysis.
➢
Moving or transporting the patient during the analysis of the heartbeat by the device may lead to a wrong or untimely
diagnosis. Reduce movements to a minimum while the heartbeat is being analysed. If the device is used in a moving
ambulance, stop the vehicle and only start driving after the shock has been delivered.
➢
In order to use the
Saver One P
, one must have completed a training course for basic or advanced cardio-pulmonary
resuscitation with the use of a defibrillator (BLS-D or ALS-D course).
➢
Avoid the use of adult defibrillation PADs (SAV-C0846) on children (ages 1-8 years or weighing 8-25kg), in this case the
Saver One P
(Ref.
SVD-B0006, SVD-B0007
) does not automatically reduce the maximum energy that can be delivered to
50J and may, therefore, become hazardous for the paediatric patient.
➢
Avoid using the universal Face to Face SAV-C0599 defibrillation PADs in adult mode on children (aged 1-8 or weighing
8-25 kg). In fact, in adult mode the
Saver One P
(Ref.
SVP-B0006-U, SVP-B0006-U-Q, SVP-B0007-U, SVP-B0007-U-Q
)
does not automatically reduce the maximum deliverable energy to 50J and therefore can become potentially dangerous
for the paediatric patient, therefore set the paediatric mode correctly from the menu if necessary.
➢
If needed, before applying the defibrillation PADs dry the patient’s chest and remove excess hair
➢
Do not allow
Saver One P
, its accessories, its parts to be dropped and/or subjected to hard impacts.
➢
Do not use damaged accessories and/or parts; otherwise, the device may be caused to malfunction.
➢
Use solely original accessories and/or spare parts.
➢
Avoid handling the device, its accessories or its parts too aggressively to avoid possible damage. Inspect the entire
system regularly.
➢
Sanitise the device in compliance with the regulations indicated in paragraph 12.3 and always make sure that the device
is switched off with the battery removed and PADs disconnected.
➢
The defibrillation pads are single-use, to be used on just one patient. Do not reuse the defibrillation pads; throw them
away after use and replace them with a new pair.
➢
Defibrillation PADs are not sterile or sterilisable.
➢
The intense or prolonged administration of cardiopulmonary resuscitation with the defibrillation electrodes applied to
the patient can damage the electrodes. Replace them if they are damaged due to use or handling.
➢
Inadequate maintenance may damage the
Saver One P
or cause it to malfunction. Comply with what is described in this User
Manual.
➢
Use non-rechargeable batteries SAV-C0903 and SAV-C0904 by A.M.I. Italia S.r.l. by the indicated expiry date.
➢
Recharge the rechargeable Li-ion battery (SAV-C0011) at least once every 4 months to ensure its perfect operation and extend its
life.
➢
Rechargeable Li-ion batteries model (SAV-C0011) must only be charged using charger model CBACCS1 (SAV-C0012) by