If the shock button is pressed, the
Saver One P
will perform the defibrillation shock. This information is highlighted by
voice (audio) and visual (colour display) prompts, shown in the table; in addition the shock button will stop flashing.
Voice prompts
Text
Shock Delivered
Shock Delivered
You can now touch the patient
The
Saver One P
performs the shock using the BTE (Biphasic Truncated Exponential) waveform with auto
compensation of the patient's thoracic impedance. The
Saver One P
shock protocol is incremental, i.e. the energy
delivered to the patient varies incrementally based on the number of shocks performed:
Saver One P 200J
: The first shock is administered at
150J
energy the following ones at
200J
Saver One P 360J
: The first shock is carried out at an energy of
200J
the second at
250J
the subsequent ones at
360J
The detected impedance value must be between 20 and 200 Ohm; if a value outside this range is detected, it is
required to position the PADs.
The shock protocol is pre-set, cannot be modified by the user and is reset at each power up. It can only be modified by
A.M.I. Italia S.r l. at the express request of the customer and endorsed by a competent body.
8.6
N
ON
-
SHOCKABLE RHYTHM
If the
Saver One P
does not detect a VF or VT during the heart rhythm analysis. This information is highlighted by voice
(audio) and visual (colour display) prompts, as shown in the following table:
Voice prompts
Text
Video Display
Shock NOT recommended
Shock NOT recommended
All rhythms other than VT and VF will be assessed as non-shockable. For more information, please see paragraphs
13.10 and 13.11.
8.7
C
HANGE OF RHYTHM
The
Saver One P
is designed to analyse the patient's heart rhythm continuously, moment by moment.
If, after recommending the shock, the device detects a sudden change in the heart rate of the patient who no longer
needs defibrillation, it will disarm automatically. This information is highlighted by voice (audio) and visual (colour
display) prompts, as shown in the following table:
Voice prompts
Text
Shock Cancelled, rhythm changed
Shock Cancelled, rhythm changed