defibrillator is able to record and store both the
SERVICE data
of the device and the
complete data
of the rescue operations
carried out. Data recording and storing is carried out automatically (cannot be disabled by
the user) both on the
internal memory
of the device and on the
memory card
when installed (with the exception of
recording voices and environmental noise). The operator can also print the data recorded directly from the device
thanks to the use of the portable thermal printer Martel MCP7830 (SAV-C1070) or the PC Saver View Express
software.
11.1
D
ATA RECORDING
The
internal memory
of the
Saver One P
makes it possible to store up to 6 hours of environmental recordings (audio),
ECG tracing, patient data (HR and Ω) and all rescue events. The stored data
can be viewed on a PC using the PC Saver
View Express software (SAV-C0019).
Two types of files are stored on the
SD Card external memory
:
•
AED1LOG.txt
This file stores all the automatic self-tests performed by the device with
their outcomes and all the
SERVICE
information.
This type of file can be viewed on a
PC via a simple reading program
•
AEDFILE.aed
This file stores the rescue data such as: environmental recordings
(audio), ECG tracing, patient data (HR and Ω) and all rescue events. This type
of file can be viewed again on a PC using the PC Saver View Express software.
The number and duration of recordings depend on the capacity of the memory card, below is an example:
Type
Capacity
Stored Data
SD Card
512 MB
Sounds, Events, Parameters, ECG.
Service
(A AEDFILE)
1,500 minutes (25 hours)
1 GB
3,000 minutes (50hours)
2 GB
6,000 minutes (100 hours)
SDHC Card
4 GB
12,000 minutes (200 hours)
The recording of voices and environmental noise of the rescue (audio) can be activated or deactivated. This procedure
is possible by activating or deactivating the device's microphone from the settings menu:
1
Turn on the device and enter the settings menu
2
Select the microphone item and set as desired
ON
Active microphone
Saver One P
makes environmental recordings
OFF
Microphone disabled
Saver One P
does not make environmental recordings
Содержание Saver One P
Страница 1: ...USER MANUAL DUAL MODE DEFIBRILLATOR SEMIAUTOMATIC MANUAL Rev 12 4...
Страница 2: ...2 QUICK START GUIDE...
Страница 76: ...76 16 CERTIFICATIONS 16 1CE CERTIFICATE...
Страница 77: ...77...
Страница 78: ...78 16 2IMQ MARK...
Страница 81: ......