Pannoramic SCAN 150 1.17 User’s Guide
4 Software Description
Multilayer Settings
By clicking the
the Multilayer Settings window is displayed. The
containing settings are the same as in Brightfield mode, but you can select extended focus
or Z-stack value for the specific channel.
Figure 55 – Filter properties window
Set the focus level value under the
Number of focus levels
box, and the
Focus step size
(one step = 0.2 μm)
.
Once you have set the correct values, tick the multilayer checkbox (EF or ZS column) of the
channel for which you want to apply these settings, or activate
Enable Multilayer
function
in the Filter Properties window.
For more information about Extended Focus, see section
4.3.5 Focus tab panel
.
July 4, 2013 - Rev. 1
3DHISTECH Ltd.
92(130)