Pannoramic SCAN 150 1.17 User’s Guide
4 Software Description
4.5.3 Preview tab
The Preview tab offers options for threshold setting and selection of the area to be scanned.
Figure 49 – Preview tab
Use threshold
•
Scan specimen with auto threshold –
This automatically detects the specimen in the
preview image.
Important!
Make sure that the detection result is acceptable for you. Otherwise set the threshold
value manually or append missing areas with the help of
Scanned
function located at the
preview toolbar. For more information see section
4.3.4 Preview area scanned by preview
camera – Preview Toolbar.
July 4, 2013 - Rev. 1
3DHISTECH Ltd.
81(130)