Pannoramic SCAN 150 1.17 User’s Guide
4 Software Description
•
Scan specimen with user set threshold
–
This sets the sensitivity of the specimen detection.
To set the desired threshold value manually, change the setting in the
Specimen threshold
level
selection box in the range of 0-255. The default value is 35.
Smaller values result in a more sensitive scanning, meaning that more areas will be digitized,
however, this may also include specks or other imperfections of the slide. Greater values
result in a less sensitive scanning, but it is possible that some of the smaller specimen areas
will not be digitized.
•
Scan area inside the marker
–
If the specimen cannot be identified as required for
digitization, (for example, weak contrast), it can help to encircle completely the specimen
with a black marker pen. You have to draw a full circle with the pen to enable the detection.
Marker pen detection will make specimen detection easier by limiting the scanned area.
If this option is selected, only the area encircled with a marker pen will be scanned. You can
define the threshold level by changing the threshold value in the
Marker threshold level
selection box in the range of 0-255.
This sets the sensitivity of the specimen detection inside the marker area. To set the desired
threshold value manually, change the setting in the
Specimen threshold level
selection box in
the range of 0-255. The default value is 35.
•
To scan everything within the marked area, check the
Scan everything within the marked
area
selection box.
Figure 23 – Scan everything within the marked are selected / not selected
General threshold options
•
Remove coverslip
–
If selected, the edges of the coverslip are not detected as specimen to be
scanned.
Depending on the threshold level, it is possible that the edges of the coverslip are detected
as areas to be digitized.
July 4, 2013 - Rev. 1
3DHISTECH Ltd.
49(130)