Pannoramic SCAN 150 1.17 User’s Guide
4 Software Description
Digital Gain
Note:
This function is camera-dependent, when working with AxioCam MRm
Rev3. the Index value ranges from 0 to 4, and in case using the
pco.edge camera, the value ranges from 0 to 6.
Enter
Index
value in the value box or set value with the arrow buttons. If you raise the
Index
value, the exposition time is halved by every step, and the image becomes noisier.
To move between the filters, click
Previous Filter
or
Next Filter
.
Multilayer Settings
By clicking the
button the Multilayer Settings window is displayed. The
containing settings are the same as in Brightfield mode, but you can select
Extended focus
or
Z-stack
value for the specific channel.
Figure 47 – Multilayer Settings window
Set the focus level value under the
Number of focus levels
box, and the
Focus step size
(one step = 0.2 µm)
.
Once you have set the correct values, tick the multilayer checkbox (
EF
or
ZS
column) of the
channel for which you want to apply these settings, or activate
Enable Multilayer
function
in the Filter Properties window.
July 4, 2013 - Rev. 1
3DHISTECH Ltd.
76(130)