Pannoramic SCAN 150 1.17 User’s Guide
6 Technical Data
Operating data
Intended site
closed room facilities
Electrical protection class
I
Internal protection degree
IP 20
Electrical safety
in compliance with DIN EN 61010-1:2001, DIN EN 61010-2-
101:2002 (IEC 61010-1:2001, IEC 61010-2-101:2002)
Over-voltage category
II
Radio interference suppression
in accordance with EN 55011 class B
Noise immunity
in accordance with EN 61326-2-6:2005
Base unit input voltage
24V DC ± 1V
Power consumption of the complete device
75 VA
24V external mains power supply unit
Intended site
closed room facilities
Electrical protection class
I
Internal protection degree
IP 20
Electrical safety
in compliance with UL 2601-1, TÜV EN 60601, CE EN60601-1-
2:2001
Over-voltage category
II
Radio interference suppression
in accordance with CISPR-11 EN 55011 class B
Line input voltage
100 V AC to 240 V AC
Line input frequency
50 / 60 Hz
Power consumption
max. 100 VA
Secondary voltage
24V DC
Secondary current
max. 4.8 A