![3DHISTECH Pannoramic SCAN 150 Скачать руководство пользователя страница 113](http://html1.mh-extra.com/html/3dhistech/pannoramic-scan-150/pannoramic-scan-150_user-manual_442163113.webp)
Pannoramic SCAN 150 1.17 User’s Guide
5 Troubleshooting and Maintenance
5.3 Maintenance
This section describes the most common maintenance solutions that can be performed by the user of
the product. For maintenance solutions not described in this manual, contact the service personnel.
5.3.1 Removing the protective covers
To replace the halogen lamp, (or to perform other maintenance work) and to interrupt a running
digitization procedure because of slide or magazine jamming, the protective covers of the
Pannoramic SCAN 150 need to be opened in order to eliminate the problem.
Before removing the covers, exit the control software, switch off the 24 V power supply and
disconnect the power cables.
Warning!
Avoid leaning over the device when opening it.
Figure 69 – Removal of covers
Five eccentric screws
(3)
are located on the bottom face of the base plate
(4)
. They attach the covers
to the base plate.
1.
Use an open-end wrench (W/F 10)
(5)
to turn the screws
(3)
on the lateral sides of the
Pannoramic SCAN 150 by 90° anticlockwise, proceeding pair-wise
(B)
. Use both hands to
take cover
(1)
carefully off upwards.
July 4, 2013 - Rev. 1
3DHISTECH Ltd.
113(130)