Replacing a Hard Drive
Generally, the only time that a hard drive should be removed from the disk array is when the
Image Server 2000 diagnostic software identifies a permanent fault. In this case, the front panel
DRIVES
LED will blink. Use the hard disk management application to identify the specific fault.
This application monitors many things about individual drives and the drive array, and will clearly
indicate which drive has failed.
Each hard drive bay is numbered 0 through 3, as are the drives. If for any reason one or more
drives have been removed, they must be reinstalled in a bay whose number matches the drive.
Installation in the wrong location will damage the data stored on all drives.
1. Turn system power OFF with the front panel button. After it has shut down, turn off the AC
power switch and unplug the AC power cord.
2. Remove the front panel as described above.
3. Unscrew the holding plate from the front of the drive sub-chassis.
4. Pull the drive being replaced from its bay about an inch (25mm). Disconnect the power
and data cables from the drive, and remove it the rest of the way.
5. Replace the drive, fasten the holding plate in place, and reattach drive cables. Note that
the data cable is numbered; re-attach it to the same drive location from which it was
removed.
6. Replace the front panel. Tighten the front panel access screws with a screwdriver.
7. Replace the AC power cord and turn on the rear panel AC switch.
8. Turn the system ON using the front panel button and check the RAID array status.
9. Check for correct operation.
NOTE: All drives in a RAID array must be identical.