Each wind measurement is calculated from the median of multiple wind samples. If
high wind speeds (above 30 m/s) are expected, more than 100 samples may be more
suitable. Contact us if you wish to change the sample setting.
E.6
Particle Settings
The sampling time in the range of 1 to 60
seconds can be specified on the Output
Settings screen. A value of 10 seconds is
recommended and is the factory default setting
for the AQSync.
K Factors for PM1, PM2.5, and PM10 are
calibration factors and can be specified by the
user.
The K-Factors are only valid at the same
site and for the same particulate type. If the local particulate source changes, the K-Factors
may no longer be valid. As shipped, the AQSync has K Factors appropriate for the Denver,
Colorado USA area.
The user should determine the appropriate K Factors for their
sampling location and adjust accordingly. See the
and in Appendix 1 for more information.
E.7
System Settings
From the Home screen, choose “Settings” and then “System Settings.” The password
set at the factory is “password”. (See Section E.7.10 for a description of how to reset
the password.)
E.7.1
Update Device
This menu enables the user to check for and download any available updates of the
AQSync User Interface, Application Programming Interface, and scripts