Alarm Message
Cause
Corrective Action
AIR TRAP FAULT
1
Air has been detected in the air
trap.
Verify that the air trap is completely
filled with saline. Inspect tubing circuit
for leaking saline. Refer to Air Entry Into
the Tubing Circuit on page 9.
If the problem persists, discontinue use
and contact your ZOLL service represen-
tative.
The air trap is full of saline but
the saline level detector indi-
cates a fault.
Completely dry the outside of the air
trap with a towel to remove any con-
densation.
Verify that the air trap is seated firmly at
the bottom of the air trap holder.
If the problem persists, discontinue use
and contact your ZOLL service represen-
tative.
COOLANT EMPTY
The coolant well is empty.
Fill with mixture of propylene glycol and
distilled water.
COOLANT LOW
The coolant level is low.
Add a 1:1 mixture of distilled water and
propylene glycol until liquid level
reaches top fill line in coolant well.
PRIMARY TEMPERATURE
PROBE DISCONNECTED
The primary patient tempera-
ture probe is disconnected.
Verify position of primary temperature
probe. Ensure that primary probe is
plugged into socket T1, or if an inter-
connect cable is used, the probe is
plugged into the cable and the cable is
plugged into T1.
PRIMARY TEMPERATURE
PROBE DISLODGED
Temperature output from the
primary patient temperature
probe changed by –0.2°C or
more within 10 seconds.
Replace probe in patient.
PUMP DOOR OPEN
The pump cover is open.
Close the cover.
PUMP FAILURE
The pump has failed.
Inspect the pump. Clear any obvious
faults and restart. If the problem per-
sists, discontinue use and contact your
ZOLL service representative.
SECONDARY TEMPERATURE
PROBE DISCONNECTED
The secondary temperature
probe is disconnected.
Verify position of secondary tempera-
ture probe. Ensure that the secondary
probe is plugged into socket T2, or if an
interconnect cable is used, the probe is
plugged into the cable and the cable is
plugged into T2.
SECONDARY TEMPERATURE
PROBE DISLODGED
Temperature output from the
secondary patient temperature
probe changed by –0.2°C or
more within 10 seconds.
Verify position of secondary temper-
ature probe.
Table 5.2. Medium priority Alarm Messages, Causes, and Corrective Actions