10
•
Inspect, and tighten if necessary, each Luer fitting (do not use instruments to tighten the fittings).
5. Inspect the tubing that returns to the pump from the patient.
6. Examine the saline bag to ensure that it has not been accidentally compromised (for example, the spike
may have damaged the bag wall).
7. Inspect the tubing from the saline bag to the saline reservoir and the pump.
Interference
If the console does cause interference with other devices, which can be determined by turning the console off
and on, try to correct the interference by one or more of the following measures:
•
Reorient or relocate the receiving device.
•
Increase the separation between the console and device.
•
Connect the console into an outlet on a circuit different from that to which the other device(s) is con-
nected.
Product Label
An identifying label is attached to the outside of the console near the power cord inlet.
The label provides safety information and identifies the manufacturer, model, serial number, power require-
ments, fuse capacity, and manufacturing date for the console.