87
TEMPERATURE PROBES DO
NOT AGREE
Temperature outputs from the
primary and secondary patient
temperature probes differ by
more than 2°C.
Determine the cause of the discrepancy
and correct it. One or both temperature
probes may need to be replaced.
SYSTEM MALFUNCTION
May be caused by a variety of
failures.
Discontinue use and contact your ZOLL
service representative.
1. Stops operation of the console.
Alarm Message
Cause
Corrective Action
DOWNLOAD ERROR. PLEASE
CHECK EXTERNAL
COMPUTER
Patient trend data download
has failed or there is no com-
puter connected to the con-
sole.
Check the connection between the
computer and the console and select
“Try again.” If problem persists, select
“Cancel” and contact ZOLL technical
support.
PATIENT DATA LOG FULL.
PLEASE DOWNLOAD NOW
Patient trend data log is full.
Only two hours of space
remain.
Download patient trend data.
If data is not downloaded, it will be
overwritten.
Table 5.3. Low priority Alarm Messages, Causes, and Corrective Actions
Table 5.2. Medium priority Alarm Messages, Causes, and Corrective Actions (continued)