Events Requiring Technical Assistance
Caution.
The console has multiple internal alarm states - see Alarms and Corrective Actions on page 85.
As described in this manual, some are remedied by the user action e.g. Filling the coldwell with coolant will
rectify the coolant level alarm.
There are others that may be alleviated, in some cases, by power cycling the console (i.e. turning it off and
then back on).
If neither user action nor power cycling the console clears the alarm –
Do not use the console.
Call a ZOLL
Service Representative for service of any alarm that does not clear.
Immediately discontinue using the console and seek advice from your ZOLL service representative if any of the
following events occur:
•
The console repeatedly trips an outlet equipped with a ground fault circuit interrupter (GFCI).
•
The protective fuse repeatedly blows.
•
Fluid is observed leaking from the console.
•
The console emits an unusual odor when operating.
•
The console produces loud or unusual noises when operating.
•
The display screen, control knob, or control buttons fail to operate as expected.
•
The console has been exposed to fire, flood, or hazardous substances.
•
The console has suffered potential damage due to improper storage, rough handling, or being dropped.