4
Voltage dips, short
interruptions, and
voltage variations on
power supply input
lines
IEC 61000-4-11
Voltage Dips >
95% reduction,
0.5 period At 0°,
45°, 90°, 135°,
180°, 225°, 270°
and 315°
Voltage Dips >
95% reduction,
0.5 period At 0°,
45°, 90°, 135°,
180°, 225°, 270°
and 315°
Mains power quality should be that of a
typical commercial or hospital environ-
ment. If you require continued operation
during power mains interruptions, we
recommend that the console be powered
by an uninterruptable power supply or a
battery.
Voltage Dips >
95% reduction, 1
period At 0°
Voltage Dips >
95% reduction, 1
period At 0°
Voltage
Interruptions >
95% reduction,
250/300 periods
Voltage
Interruptions >
95% reduction,
250/300 periods
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical loca-
tion in a typical commercial or hospital
environment.
The Thermogard XP console is intended for use in the electromagnetic environment specified below. The
customer or user of the console should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic environment
guidance
Conducted RF
immunity
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
150 kHz to 80 MHz
Portable and mobile RF communications
equipment should be used no closer to
any part of the console, including cables,
than the recommended separation dis-
tance calculated from the equation appli-
cable to the frequency of the transmitter.
Recommended separation distance
150 kHz to 80 MHz
d = 1.2
, 80 to 800 MHz
d = 2.3
, 800 MHz to 2.7 GHz
6 Vrms in ISM
bands within
150 kHz and
80 MHz
6 Vrms in ISM
bands within
150 kHz and
80 MHz (see Note
3)
80% AM at 1 kHz
80% AM at 1 kHz
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic environment --
guidance
Table 1.2. Guidance and Manufacturer’s Declaration – Electromagnetic Immunity