Figure 4.16. Standby Timer Menu
If the console has been left in Standby for more than the specified time, an alarm will sound to remind that the
console remains in Standby. Pressing the knob will reset the timer. The Standby Timer function will continue
until either:
•
The console is placed into Run.
•
The Standby Timer Menu is used to deactivate the Standby Timer.
The Standby Timer alert may be silenced for two (2) minutes by pressing the mute button.
Figure 4.17. Standby Timer Alert
T1/T2 Behavior
WARNING.
Dislodged Foley Catheter Temperature Probe. The console can detect when a patient temperature
probe is dislodged suddenly from the patient. However, it is possible for the Foley catheter patient temperature
probe (T1) to become dislodged from the bladder to rest on the perineum or within the fold of the thighs. In
this position, the console may not detect the dislodgement from the patient and will underestimate the