Temperature Probe Connectors
The front of the console features two connectors, labeled “T1” and “T2” which are used for connection to
patient temperature probes. The primary patient temperature probe is plugged into connector T1. The second-
ary probe is plugged into connector T2.
Figure 2.4. Temperature Probe Connectors.
Pump
Sterile saline solution is circulated through the heat exchange coil and the catheter by a high-performance,
compact roller pump. It pumps by peristaltic action on the tubing installed in the pump head. The pump rota-
tion speed is accurately controlled by an electronic speed control system.
If a tubing leak or failure occurs in the pump raceway, the saline solution will cause corrosion in the moving
parts of the rotor. Contact your ZOLL service representative.
WARNING. Finger injuries.
Be careful when inserting the pump tubing that you do not
catch your fingers with the roller. When the console is operating, do not attempt to circum-
vent the safety interlocks on the peristaltic pump lid. Do not place fingers or foreign objects
into the pump raceway when the pump is turning. The peristaltic pump has sufficient
torque to severely damage a finger.