Alarms
5-14
www.zoll.com
906-0330-01-01 Rev. D
Pop Up Messages
Notifications
Pop Up Name
Information
Message
Silent/Dark Mode Enabled
Pop up triggers when you attempt to
begin operation using Silent/Dark
mode. In order to proceed, you must
press the
Confirm button
to begin
operation.
Note: Not all devices are provided with
the capability to operate with Silent/
Dark mode, please check with your
organization.
Press Accept Key to
Enter Silent/Dark Mode
Now
Smart Flow Cal Enabled
Pop up triggers when you attempt to
begin a calibration of the Smart Flow
feature. The pop up instructs you to
ensure that all accessories are
connected to the device but not
attached to a patient. In order to
proceed, you must press the
Confirm
button
to begin the calibration
procedure.
Attach accessories to
aspirator before starting
Smart Flow calibration.
Press the Accept key to
begin calibration.
Smart Flow in Progress
Pop up triggers after you press the
Smart Flow Cal: button and the Smart
Flow calibration process begins.
The pop up tells you that the
calibration is in progress and to wait
for the completion of the calibration.
The pop up clears automatically
when the calibration process
completes.
Please wait until
calibration is complete.
This pop up clears and
the previous operation
procedure resumes when
calibration is complete.
Notification
Information
Message
Altitude Conflict/Limit
Notification occurs when the vacuum
setting is within 76% of environmental
(barometric) pressure.
Target too high for
ambient pressure.