Measuring system XS22-E2010
9 Use of ZrO
2
-probes
21
XS22_E2010
9 Details for the use of ZrO
2
-probes in vacuum technology
9.1 Theoretical relations
Potentiometric ZrO
2
-probes measure the difference of the oxygen partial
pressure at both electrodes.
One electrode is fixed inside, the other one outside the ZrO
2
-tube. Air as
reference gas flows around the inner electrode.
The NERNST equation applies:
)
(
)
(
ln
4
2
2
gas
measuring
p
air
p
F
RT
U
O
O
=
(Nernst equation)
Condition: isothermal electrodes
U
– cell voltage
R
– molar gas constant, R = 8.31 J/(mol
.
K)
T
– Kelvin temperature
F
– Faraday constant, F = 9.64
.
10
4
C/mol
p
O2
(air) – partial pressure of oxygen at reference electrode
p
O2
(measuring gas) – partial pressure of oxygen at measuring electrode
After inserting the known values of the natural constants the following applies:
]
/
/
42
.
46
exp[
)
(
)
(
2
2
K
T
mV
U
gas
reference
p
gas
measuring
p
O
O
⋅
−
⋅
=
]
/
/
42
.
46
exp[
2064
.
0
)
(
2
K
T
mV
U
p
gas
measuring
p
L
O
⋅
−
⋅
⋅
=
0.2064 refers to the oxygen concentration in air at 50 % relative humidity and
p
L
is the air pressure. The pressure unit of p
O2
(measuring gas) and p
L
can
be chosen user-defined but must be identical for both pressures. The
numerical value in this equation requires the probe signals U used in mV and
T in Kelvin.
For the calculation of the oxygen concentration
ϕ
O2
in vacuum, which is the
ratio of the number of oxygen molecules to the total of gas particles, the total
pressure in the vacuum chamber is necessary.
Summary of Contents for XS22
Page 2: ...2 XS22_E2010...
Page 4: ...4 XS22_E2010...
Page 20: ...20 XS22_E2010...
Page 24: ...24 XS22_E2010...
Page 26: ...26 XS22_E2010...