Measuring system XS22-E2010
8 Fault causes and clearance
19
XS22_E2010
8 Potential fault causes and fault clearance
The following list presents an overview of potential fault appearances. Faults not in the list
require a consultation with the manufacturer.
Symptoms
Fault causes
Fault clearance
Red LED off (no supply
voltage)
No supply voltage at the
probe
Check power supply
Internal fuse switched off
Send probe to service
Green LED off (no ready-
signal)
Probe is too cooled by gas
flow
Install probe flow-protected
Heater of the probe defective Send probe to service
Thermocouple defective
Send probe to service
Output signal > range
Current / voltage transformer
for cell voltage is overdriven
Convert amplifier to higher
measuring range 1200 mV