Measuring system XS22-E2010
4 Composition
12
XS22_E2010
Digital output:
………… RS 232-Interface (9600 Baud)
Pin assignment RS 232
The RS232 interface must be connected with a computer by SUB-D-cable
(9pol., 1:1, uncrossed)!
Transfer rate: max. 9600 Baud, adjustable
Protocol of the serial interface (CR = carriage return)
Set
Feedback signal
(example)
Transferred
measuring value
Parameter
M2CR
M2x.xxExxCR
M22.06E+05
2.06*10
5
ppm O
2
Oxygen concentration in
ppm
A1CR
A1xxx.xCR
A120.9
20.9 mV
Cell voltage in mV
A2CR
A2xxx.xCR
A2749.9
749.9 °C
Measuring temperature
in °C
The parameters are set by the keys of the front cover (see chapter 4.1).
PC-software for display and storage of the measuring values is available on request
(extra charge).
Error messages via RS232
ERROR0
Transfer error RS232 (or wrong command)
ERROR1
Warm-up (cell temperature too low, < 30 min)
ERROR2
Cell temperature too low (< set temp. – 10 °C, > 30 min)
ERROR3
Thermocouple broken
ERROR6
System error
Stop bits
1
Parity
no
Data bits
8
Handshake
without
MD
TXD
RXD