49
CHAPTER
TECHNICAL DATA
ZIRKONZAHN – OPERATION AND MAINTENANCE MANUAL M1 MILLING UNIT COMPACT LINE – VERSION: 13/05/2021
CONNECTION TO THE POWER SUPPLY SYSTEM VIA RESIDUAL-CURRENT CIRCUIT BREAKER
-
6JGOKNNKPIWPKVUCTGGSWKRRGFYKVJCPKPVGITCVGF'/%OCKPUſNVGTVQOGGVVJGEWTTGPV'/%
standard 2014/30 EC.
-
&WGVQVJGRTKPEKRNGWUGFVJG'/%OCKPUſNVGTRTQFWEGUNGCMCIGEWTTGPVVQITQWPFXKC2'
- Provide a suitable alternative protective measure.
- Permissible residual-current circuit breakers, Type A or B (30 mA).
- The use of residual-current circuit breakers, Type AC, is prohibited!
-
6JGTGNGCUGXKCCTGUKFWCNEWTTGPVEKTEWKVDTGCMGTCVVJGſTUVUVCTVWRKUIGPGTCNN[CHCWNVQHVJG
power connection, not of the machine.
ENVIRONMENT AND STORAGE CONDITIONS
- Degree of contamination according to IEC 61010-1:2.
- Up to 2000 m above sea level.
- Ambient temperature, in the range from 15 °C to 30 °C.
- Do not expose to direct solar radiation.
- Maximum relative air humidity of 75% at 30 °C;
- Do not place it near an air conditioner. Avoid cold draughts.
PERMISSIBLE TRANSPORT CONDITIONS
The following conditions must be observed when transporting the furnace:
- Keep it in a dry place.
- Do not place it in a cold environment.
- Protect it from dusts.
- Do not place it near corrosive agents (e.g. salt water).