Error messages
Troubleshooting
Disposal
13
Page 23
Error message
Handpiece temperature
The handpiece temperature is not properly regulated. Please change the handpiece.
If treatment is being performed using two handpieces, treatment can be continued
using the intact handpiece.
Error code (left/right):
0: normal
1: error
Error message
Water pump
The water pump is not functioning properly.
Please contact customer services.
Error message
Programme reset
The device is not functioning properly and needs to be turned off. If, when switched
on again, the error message continues to appear, please contact customer services.
Error message
Communication
There is a communication error between the CPU and the LCD.
Please contact customer services.
Customer services
If frequent malfunctions occur or if they cannot be resolved as described above,
please inform customer services.
You can reach the service department via your sales representative or by calling our
Neu-Ulm office.