Device setup
5.2 Setup the device
5
Page 7
Pluging in the power
cable
Connect the power cable to the corresponding socket (6) on the device and connect
the cable to the mains.
Note:
The device may only be connected to earthed sockets.
Note:
All cables must be protected against jamming or other mechanical damage.
Connect the
applicators
Insert the applicator into its holder. Then connect to the device by placing the
connector into the socket until you hear a clicking sound twice.
To remove the applicator, press the buttons on both sides of the connector
simultaneously.
Turn off the device and disconnect the power cable before changing the handpiece.
Warning
!
If the applicators are not properly connected to the unit, it may cause water leakage.
Fill with distilled water
Open the door at the back of the unit by moving the black slider to the right. Open
the filler pipe by switching the white “Fill”-cap a left quarter-turn and pull it off.
Connect the filling hose with the filler neck and let the distilled water run slowly until
the red mark is reached.
Warning
!
Warning
!
Do not use tap water, only use distilled or deionised water.
The water should be replaced every 3 months.