Error messages
Troubleshooting
Disposal
13
Page 22
Error message
Lack of water
first warning
If this symbols appears on the left side of the screen bar, there is a shortage of water
in the unit. Please fill with distilled or deionised water as described in chapter 5.
second warning
If, after the first warning (see above), no water is supplied, this symbol appears after
24 hours of operation. Please fill with distilled or deionised water as described in
chapter 5.
If there is less than 900ml of water in the unit, safe operation is no longer possible
and the error message appears. Please fill with distilled or deionised water as
described in chapter 5.
Error message
Handpiece not
connected
Handpiece not or improperly connected.
Make sure that the handpiece is properly connected. The plug
must be fully engaged (audible through a double click).
Error message
Handpiece sensor
The handpiece's temperature sensor does not work properly. Please change the
handpiece.
If treatment is being performed using two handpieces, treatment can be continued
using the intact handpiece.