Cleaning
Disinfection
9
Page 16
Note:
Always keep the device and the handpieces clean in order to obtain the best
possible results.
Before any maintenance or cleaning, the device must be switched off with the main
switch and the power cable must be unplugged.
Cleaning the unit during operation can be dangerous for the user and/or device.
Neither the unit nor the handpieces may be immersed in fluids.
Make sure that no liquids penetrate the unit during cleaning and disinfection. Do not
use sprays.
If liquid does penetrate the device or applicator during cleaning or disinfection,
please decommission the unit, protect it against recommissioning, and contact the
service department.
When cleaning and disinfecting, take care not to damage any markings on the
device (e.g. warnings and labels).
Note:
If 70% isopropyl alcohol is used for cleaning, the device must be completely dry
before it is used.
The device and its applicator are not considered critical in relation to hygiene when
used on non-injured and healthy skin.
Applicators
Use a soft cloth for cleaning and disinfection.
After each application, the applicator's inner walls must be cleaned with a damp
cloth.