Device setup
5.2 Setup the device
5
Page 8
During the first installation, the entire system needs to be filled with distilled water.
Start the unit for 30 seconds (without selecting a treatment) and then fill again up to
the red mark.
Warning
!
This procedure must be repeated for each handpiece when connected for the first
time.
Switch the device on
Turn on the device with the power switch (6).
Finally, perform a function test (see chapter 12).
Note:
Check the following before each use:
•
Are the housing and handpieces clean and sanitised?
•
Are the unit's wheels secure on the ground?
•
Is the power cord plugged into an earthed socket?
Note:
When starting the unit, the handpieces should be connected and in their holders.
Switch off the unit.
The power is turned off via the power switch (6).
Remove the power cord to disconnect the device completely (all poles) from the
mains.
Warning!
If the unit is not used for a long time or is exposed to colder temperatures (e.g.
during transport), it is advisable to remove the water through the drain cock.