Application information
2
Page 2
General application
notes
Z
Lipo
has been developed in accordance with international safety standards.
Before using Z
Lipo
on the patient, the user/technical personnel should familiarise
themselves with the instructions for use, the individual treatments as well as the
application areas/contraindications, warnings and treatment advice. Additional
sources of information about the treatment should also be observed.
The device may only be used by physicians or qualified personnel. Make sure that
all users are familiar with the use of the equipment and know how to turn the power
off immediately in an emergency.
Warning
!
During use, the device must be located in a position that allows direct access to the
central power supply so that this can be disconnected at any time.
Only personnel who have been informed of the safety procedures should have
access to the treatment room.
Never leave the unit switched on, open or unattended.
Warning
!
Do not use the device if it has been dropped or if has been struck by any other
heavy force. Try to keep shocks and vibrations to a minimum.
Warning
!
Ensure that the handpieces do not come into contact with hard materials which may
damage the handpieces.
Warning
!
Before use, ensure that the device is operated via a properly grounded socket
(electrical installation in accordance with DIN VDE 0100 Part 710). The device may
only be operated with the power cable supplied. The power cable must be protected
against mechanical stress.
Warning
!
Z
Lipo
is not suitable for operation in areas with explosive, flammable or combustive
atmospheres. Corrosive materials can damage the device. Dust and hair should be
reduced to a minimum.
Warning
!
To prevent the risk of electric shock; the device's main plug must be disconnected
from the mains supply before performing any maintenance or cleaning activities.
Warning
!
Only use accessories from Zimmer MedizinSysteme GmbH.
Warning
!
Operation of the device may interfere with the strong electromagnetic fields of other
devices. Please maintain a safe distance of at least 2 metres.
Warning
!
To ensure adequate ventilation, the device must be at least 0.5 metres from other
devices or walls.