Basic settings
6
Page 11
Adjusting the phases
(in Auto-mode)
By selecting one of the three phases, a selection window opens in which the phases
can be defined.
The structure and first phase setting will be explained below.
The other two phases are constructed analogously.
1 2
(1) Pulsing
The vacuum can be selected unpulsed or with 3 different pulses.
The selected pulsing appears in orange.
(2) Temperature
The applicator temperature can be selected between -10°C and 5°C. By pressing
the arrow keys the temperature can be increased or decreased.
(3) Treatment time
The treatment time is adjusted by pressing the arrow keys.
Note:
The treatment time is limited to a maximum of 60 minutes. This includes the sum of
treatment times of all individual phases.
(4) Reset
The selected parameters will be discarded by pressing the button, upon which the
previous settings will be displayed.
(5) Save
The selected parameters will be used for treatment.
Note:
The selected parameters also remain after being saved using the current treatment
session. When the unit is next powered on, the parameters from the last session are
loaded automatically.
(6) Cancel
The menu is exited without saving.