Zimmer Z Lipo Instructions For Use Manual Download Page 1

Instructions for Use

Z

Lipo

EN

Summary of Contents for Z Lipo

Page 1: ...Instructions for Use ZLipo EN ...

Page 2: ......

Page 3: ...Figure Front panel Fig 1 1 Display 2 Applicators 1 2 ...

Page 4: ...Figure Rear panel Fig 2 3 Applicator connections 4 Door release 5 Fuses 6 Power switch and socket for power cable 3 4 5 6 ...

Page 5: ...Figure Applicators Fig 3 7 Applicator head 8 Applicator hose 9 Applicator connector Fig 4 applicators available in 4 sizes 7 9 8 ...

Page 6: ...uctions for use this symbol indicates danger of possible damage to the device Applied part type BF Follow instructions for use Products marked with the adjacent symbol must not be disposed of with household waste Serial number Item number Manufacturer Marking in accordance with EU Low Voltage Directive ...

Page 7: ... symbols Page 1 Application areas Contraindications Side effects 1 2 Application information 2 3 Warnings 4 4 ZLipo in brief 5 5 Device setup 5 1 Preparation of the site 5 2 Setup the device 6 7 6 Basic settings 9 7 Operating instructions 14 8 Technical information 15 ...

Page 8: ...Function test 21 13 Error messages troubleshooting disposal 22 14 Certificate of conformity 25 Valid for the ZLipo device These operating instructions are an integral part of the device They must be stored with the device and be accessible at all times to anyone authorised to operate the device The instruction manual is valid from September 2018 ...

Page 9: ...ommended in case of Pregnancy nursing mothers Cardiovascular disease Liver disease Kidney disease Malignant tumours Diabetes Infections eczema or acute inflammation in the treatment area Hypertrophic scars in the treatment area Patients with pacemakers Constant intake of anti inflammatory drugs Skin diseases in the treatment area Redness Treatments on the chin are not recommended in case of Untrea...

Page 10: ... if it has been dropped or if has been struck by any other heavy force Try to keep shocks and vibrations to a minimum Warning Ensure that the handpieces do not come into contact with hard materials which may damage the handpieces Warning Before use ensure that the device is operated via a properly grounded socket electrical installation in accordance with DIN VDE 0100 Part 710 The device may only ...

Page 11: ...e patient s fat and skin condition The automatic mode is preset with proven specifications However the user can adjust the settings to the patient s condition The treatment parameters can be defined individually The Auto mode is preset with treatment recommendations Depending on the patient s condition these parameters can be adjusted The treatment results may vary The cooling fleece is essential ...

Page 12: ...uctions pertaining to the site duration and intensity of treatment require medical knowledge and may only be given by licensed physicians therapists and members of the medical auxiliary professions or other qualified persons These instructions must be followed without fail Before plugging in and removing the applicator the device must be switched off Do not touch the handpiece connectors during op...

Page 13: ...cells are phased out during normal metabolism within approximately 90 days Why use ZLipo The simultaneous use of two applicators reduces treatment times The use of differently sized applicators enables efficient and effective treatment in nearly all areas of the body The modern clearly understandable colour display shows all parameters relevant to the treatment and the modern touchscreen controls ...

Page 14: ...m the wall or other airflow The device may affect your computer or other electrical or medical equipment It should be kept at least 2 metres away from such devices The device can be plugged into an outlet with protective earth contact Pay attention to the data indicated on the label on the power supply The device should not share the power line to devices with high power usage such as air conditio...

Page 15: ...ove the applicator press the buttons on both sides of the connector simultaneously Turn off the device and disconnect the power cable before changing the handpiece Warning If the applicators are not properly connected to the unit it may cause water leakage Fill with distilled water Open the door at the back of the unit by moving the black slider to the right Open the filler pipe by switching the w...

Page 16: ...nction test see chapter 12 Note Check the following before each use Are the housing and handpieces clean and sanitised Are the unit s wheels secure on the ground Is the power cord plugged into an earthed socket Note When starting the unit the handpieces should be connected and in their holders Switch off the unit The power is turned off via the power switch 6 Remove the power cord to disconnect th...

Page 17: ...h dual applicators can be selected Note Treatment can depending on the treatment area be performed with just one applicator left or right or with two applicators simultaneously Note The programme specifications and any changes are identical for each of the 3 possible choices so the steps below are only explained with reference to the treatment with the left applicator The selected applicators left...

Page 18: ... intensity 1 10 The default is an intensity of 5 5 Operating mode The operating mode Auto or Manual can be selected by pressing the appropriate key and will then appear in orange 6 Programme Displays the programme sequences 7 Phase Setting Selects the menu for changing the programme sequences 8 Treatment start Pressing the start button begins the treatment 9 Tones Turns the beeps on and off when p...

Page 19: ...row keys the temperature can be increased or decreased 3 Treatment time The treatment time is adjusted by pressing the arrow keys Note The treatment time is limited to a maximum of 60 minutes This includes the sum of treatment times of all individual phases 4 Reset The selected parameters will be discarded by pressing the button upon which the previous settings will be displayed 5 Save The selecte...

Page 20: ...arrow keys the vacuum can be increased and decreased Vacuum intensity 1 10 The default is an intensity of 5 5 Operating mode The operating mode Auto or Manual can be selected by pressing the appropriate key and will then appear in orange 6 Parameter Displays the parameter 7 Treatment start Pressing the start button begins the treatment 8 Tones Turns the beeps on and off when pressed 9 Colour of Di...

Page 21: ...nd 5 C By pressing the arrow keys the temperature can be increased or decreased 2 Treatment time The treatment time is adjusted by pressing the arrow keys Note The treatment time is limited to a maximum of 60 minutes 3 Pulses The vacuum can be selected unpulsed or with 3 different pulses The selected pulsing appears in orange 1 2 3 ...

Page 22: ... via the start button on the screen or the start button located on the applicator During application patients should not be left alone They should be checked at regular intervals at least every 15 minutes Please monitor the temperature on the screen for 2 3 minutes until the ideal temperature has been reached If the ideal temperature does not remain constant please stop the treatment and call the ...

Page 23: ...at 700 to 1060 hPa Storage Transport without filled water 10 to 50 C 10 to 90 relative humidity without condensation at 500 to 1060 hPa Applicators Dimensions Medium handpiece W 108 mm D 188 mm H 138 mm Large handpiece optional W 118 mm D 220 mm H 142 mm Small handpiece optional W 108 mm D 172 mm H 120 mm Mini handpiece optional W 64 mm D 113 mm H 99 mm Note Storage and transport only in original ...

Page 24: ... Do not use sprays If liquid does penetrate the device or applicator during cleaning or disinfection please decommission the unit protect it against recommissioning and contact the service department When cleaning and disinfecting take care not to damage any markings on the device e g warnings and labels Note If 70 isopropyl alcohol is used for cleaning the device must be completely dry before it ...

Page 25: ...cially available alcohol free disinfectant for metal and plastics that has a bactericidal virucidal and fungicidal effect Observe the application instructions provided by the manufacturer Wipe all surfaces with a soft cloth soaked in the cleansing agent but not dripping or pre saturated wipes according to the manufacturer s instructions Follow any instructions for drying or post cleaning LCD Touch...

Page 26: ...ized applicators 58000500 20 Cooling fleeces 117 1 Power cable 10102366 1 Instructions Accessories Art No 58000400 1 Small applicator 58000410 1 Medium sized applicator 58000420 1 Large applicator 58000559 1 Mini applicator 58000510 1 Pack of 25 cooling fleeces 58000560 1 Pack of 50 cooling fleeces mini ...

Page 27: ...ammable or combustible materials Water and electricity are a dangerous combination Do not use ZLipo in a wet environment Do not dip any component directly into liquid The use of external cables and accessories may result in increased emissions or decreased safety Foreign accessories may not be connected to the device Do not use a damaged cable As manufacturer Zimmer MedizinSysteme GmbH shall only ...

Page 28: ...at the back of the unit by moving the black slider to the right To release the air filter slide the black slider down and move the air filter to the left towards the open lock Then pull the air filter down Clean the air filter with a mild soap and dry it afterwards Then put the air filter back in and turn it to the right towards the closed lock ...

Page 29: ...d the mains connection is established Function test Turn on the device Choose an applicator that is connected to the device Press the start button on the screen The LEDs on the applicator are steady blue and the cooling starts The animation will start on screen and the time counts down Note Turn off ZLipo after completion of the function test If an application should be carried out after the funct...

Page 30: ...ter as described in chapter 5 If there is less than 900ml of water in the unit safe operation is no longer possible and the error message appears Please fill with distilled or deionised water as described in chapter 5 Error message Handpiece not connected Handpiece not or improperly connected Make sure that the handpiece is properly connected The plug must be fully engaged audible through a double...

Page 31: ...rly Please contact customer services Error message Programme reset The device is not functioning properly and needs to be turned off If when switched on again the error message continues to appear please contact customer services Error message Communication There is a communication error between the CPU and the LCD Please contact customer services Customer services If frequent malfunctions occur o...

Page 32: ...many Tel 49 731 9761 0 Fax 49 731 9761 118 info zimmer de www zimmer de Disposal The device may only be returned to the factory in its original packaging It must only be disposed of via the factory in Neu Ulm In foreign European countries please refer to national regulations for disposal Contact your distributor if necessary ...

Page 33: ... Directive LVD Official Journal of the European Community 29 03 2014 L 96 357 EN 60335 2 23 2003 A1 A11 AC 2012 EN 60335 1 2012 Directive 2014 30 EU EMC Directive Official Journal of the European Community 29 03 2014 L 96 79 EN 62233 2008 Directive 2011 65 EU RoHS Official Journal of the European Community 01 07 2011 L 174 88 The device bears the CE mark This declaration is valid from 16 May 2016 ...

Page 34: ......

Page 35: ......

Page 36: ...nSysteme GmbH Junkersstraße 9 89231 Neu Ulm Germany Tel 49 7 31 97 61 291 Fax 49 7 31 97 61 299 export zimmer de www zimmer aesthetics com Instructions for Use EN 10 102 366 BäZi 0918 I Version 7 I Subject to change ...