Infinity
®
Orange
™
Operator’s Manual
36
9. ADDITIONAL TECHNICAL INFORMATION
Electromagnetic Emissions/Interference
Electromagnetic emissions may affect the operation of any electronic medical de
-
vice, including enteral feeding pumps.
The Infinity Orange will not be affected by electromagnetic emissions in most
environments. However, some electromagnetic fields produced by personal com
-
munication equipment, household appliances, or occupational tools may cause
electromagnetic interference (EMI) which can affect the pump.
Possible sources of electromagnetic interference with electronic medical devices
include, but are not limited to: cellular phones, cordless telephones, microwave
ovens, anti-theft/security systems, blenders, and high-powered tools (i.e. drills,
saws, chain saws). If electromagnetic emitting devices are operated within one
yard/meter of the Infinity Orange, the pump may automatically shut off and settings
may return to their default values. Check the pump regularly if operating near
sources of electromagnetic emissions.
The Infinity Orange can safely be operated on commercial aircraft and is designed
in accordance with EN 60601-1-2, EN 60601-1-4 and RTCA DO160D standards
for electromagnetic emissions and immunity.
ADDITIONAL TECHNICAL INFORMATION
Guidance and manufacturer’s declaration - electromagnetic emissions
The Infinity Orange is intended for use in the electromagnetic environment specified
below. The customer or the user of the Infinity Orange should assure that it is used in
such an environment.
Emissions Test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Infinity Orange uses energy only
for its internal function. Therefore, its
RF emissions are very low and are not
likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
The Infinity Orange is suitable for use in
all establishments, including domestic
establishments and those directly con
-
nected to the public low-voltage power
supply network that supplies buildings
used for domestic purposes.
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies