2. DIRECTIONS FOR USE
© Copyright 2008 ZEVEX Incorporated
11
Step 6
Confirm food type to be delivered (formula or human milk) before starting pump
by pressing the
FOOD TYPE
key. Press the
+
or
-
key to change between formula
and human milk.
Step 7
Connect barbed adapter to the patient’s enteral feeding tube.
Step 8
Press the
RUN/PAUSE
key. Pump will begin run
-
ning. Display will show programmed rate and the arcs
around the run symbol will rotate (
Figure 2-13
).
While pump is running, the following may be viewed
by pressing the appropriate key: rate, dose, and food
type.
To view the amount delivered in the current feeding,
press the
VOL/TOTAL
key (
Figure 2-14
). This counter
will reset to 0 when the previously programmed dose
has been completed and a new feeding is started.
This counter will also reset to 0 if a feeding is inter
-
rupted and rate or dose is changed, or if the pump is
turned off and then back on again.
To view the amount delivered over the course of sev
-
eral feedings press the
VOL/TOTAL
key a second
time. Display will read TOTL then the amount. This
counter never resets itself. It can only be reset by the
user (
Figure 2-15
).
While pump is running, the settings cannot be changed and the Prime feature is
disabled.
If you want to stop pump at any time, press the
RUN/PAUSE
key. Display will show
rate, and pump will be in pause mode. Pump will save the memory of where it was in
the feeding before being paused. Or, turn pump off by pressing the
ON/OFF
key.
Figure 2-15 Total Amount
Delivered
Figure 2-13 Pump is
Running
Figure 2-14 Amount
Currently Delivered