2. DIRECTIONS FOR USE
© Copyright 2008 ZEVEX Incorporated
7
2. DIRECTIONS FOR USE
Recommendation for First Use:
Since battery may not be fully charged when pump is first received, it is recom
-
mended that battery be charged for 6 hours prior to operating on battery power
(
see Page 15 for additional information
).
Priming and Loading the Disposable Set
Step 1:
4
If you are using the spike set with a vented
bottle or pre-filled bag:
Remove protective cover from spike, and
insert spike into spike adapter of formula con
-
tainer (
Figure 2-1
).
4
If you are using the 100 ml, 500 ml, or 1200
ml disposable set:
Hold bag upright and pour in feeding solution
(
Figure 2-2
). Close cap securely.
NOTE: Blenderized or aggressively mixed solutions
may have foam. If using this type of solution, allow
it to sit for 10 to 15 minutes before pouring into bag.
This will reduce the chance of an alarm due to air
in the tubing.
Step 2:
Remove protective cover from barbed adapter. If
using pole clamp, protective cover may be inserted
into the receptacle on back of clamp (Figure 2-3).
NOTE: If set is to be used with a carry pack, all air
must be removed from bag and tubing.
Continue
to step 3 for instructions on removing air
. If set
is to be hung above pump, i.e. on an IV pole,
you
may skip to step 4
.
Figure 2-1 Spike Container
Figure 2-2 Fill Bag
Figure 2-3 Remove Cover from
Barbed Adapter / Place Cover
on Pole Clamp
Cover Receptacle