Operator’s Manual
20
3. ALARMS, MESSAGES, AND INDICATIONS
Alarms
ER01 - ER99 (Figure 3-1)
Pump will sound a continuous beep and display will
read ER followed by a two digit number to indicate a
self-test has failed.
To clear alarm:
Turn the pump off and then back on.
WARNING:
If an error occurs (ER01 - ER99 will ap
-
pear in display), all settings and volume counters
should be checked and verified before starting a feed
-
ing.
LOAD SET (Figure 3-2)
Pump will sound a dual tone beep repeatedly and
display will read LOAD SET.
Why:
This alarm will sound when pump attempted
to run with the set improperly loaded or missing.
To correct the problem:
Press the
RUN/PAUSE
key to silence alarm and place pump in pause
mode, then check the following:
1. An Infinity brand disposable set is loaded into
pump properly and door is closed.
2. Check cassette receptacle for cleanliness, es
-
pecially around pressure sensors. If cleaning is
necessary, refer to
CHAPTER 5 - CLEANING
(Page 29)
.
LOW BATT (Figure 3-3)
When 1 hour or less of battery life is available, display
will flash LOW BATT every 3 seconds and pump will
beep every 2 seconds to indicate battery is low and
pump will soon stop running.
To correct the problem:
Plug in AC Adapter/Char
-
ger to recharge battery.
Figure 3-2
LOAD SET Alarm
Figure 3-3
LOW BATT Alarm
Figure 3-1
ER01 - ER99 Alarm