2. DIRECTIONS FOR USE
© Copyright 2008 ZEVEX Incorporated
9
Step 5:
Loop silicone tubing around pump wheel stretching
lightly. Seat cassette into pump (
Figure 2-8
).
Close
pump door
(
Figure 2-9
).
NOTE: If 1200 ml set is to be used with Mini Back
-
pack, load set into back section of pack and thread
cassette through port in bottom corner of pack prior
to loading cassette into pump.
Step 6:
Press and hold
ON/OFF
key for 1.5 seconds to
turn pump on. While pump runs through a self test,
display will light and an audible alarm will sound
as pump displays the nine digit serial number three
digits at a time for one second each. The display
will then show the letter ‘N’ followed by a number,
which is the software revision.
Next, all segments of display will be shown for 2
seconds.
Verify all display segments and sym
-
bols are active.
The self test is complete and pump will then display
last programmed rate and will be in pause mode.
If any air is still in the tubing, continue to Step 7 to
use pump prime feature.
Step 7:
Press and hold
PRIME
key. Alarm will sound once
and pump will begin pumping at the maximum rate of 332 ml/hr. Display will read
“TO STOP LET GO” (
Figure 2-10
). Once all air is removed from tubing, release
key. Pump will stop, display will revert to last programmed rate, and pump will be
in pause mode.
Figure 2-8 Seat Cassette
Figure 2-9 Close Pump Door
Figure 2-10 Priming Indication