5. CLEANING
© Copyright 2008 ZEVEX Incorporated
29
CLEANING
WARNING:
To avoid electrical shock, never clean pump or Infinity Orange AC
Adapter/Charger with charger plugged into an outlet or pump turned on.
WARNING:
Make sure the Infinity Orange AC Adapter/Charger is completely dry
before plugging into an electrical outlet.
To clean the Infinity Orange enteral feeding pump:
Pump may be cleaned with warm soapy water (standard dish soap) and a non
-
abrasive sponge or soft cloth. Use a cotton swab to clean pathways of cassette
receptacle and to remove orange silicon residue from pump wheel roller pins.
OR
- Apply one of the following solutions for ap
-
proximately 10 minutes, then wipe pump clean with
a damp cloth or sponge:
4
5% bleach and water solution
4
Multipurpose household disinfectant cleaner
Rinse pump by holding under a stream of warm
water. Then, dry with a clean cloth.
NOTE: Avoid harsh cleaners/disinfectants. They may
cause damage to pump surfaces and may affect pump
functionality. ZEVEX has tested the following disinfectants for acceptability for cleaning/
disinfecting the Infinity Orange pump. These cleaning products are categorized below:
Acceptable
Unacceptable
Metrex CavaCide
Ball Ruthless
MetriCide 28
Wex-Cide 128
5% Bleach and Water Solution
Alcohol
Warm Soapy Water (Standard Dish Soap)
Vesphene Ilse
To clean the Carrying Packs:
The Infinity brand packs are machine-washable. Use cold water and gentle cycle,
whenever possible. Hang to dry.
To clean the AC Adapter/Charger:
Infinity Orange AC Adapter/Charger normally does not require cleaning. When
desired, a dry or slightly damp cloth may be used to clean the outside surface of
AC Adapter/Charger while it is disconnected from the wall outlet.
Figure 5-1 Rinsing Pump
Under Stream of Water