2. DIRECTIONS FOR USE
© Copyright 2008 ZEVEX Incorporated
17
Max Rate:
Rate range is 0.1 ml/hr up to 300 ml/hr. However,
the rate may be programmed to limit how high the
rate can be set. The limits are 50, 100, 150, 200,
and 250 ml/hr. Rate is not limited if user preference
setting is programmed to NO MAX. Pump default is
NO MAX (
Figure 2-23
).
Max Dose:
Dose range is 0.1 ml up to 3000 ml. However, the
dose may be programmed to limit how high the
dose can be set. The limits are 10, 20, 30, 40, 50,
60, 70, 80, 90, 100, and 120 ml. Dose is not lim
-
ited if use preference setting is programmed to NO
MAX. Pump default is NO MAX (
Figure 2-24
).
To
adjust these settings:
Key Sequence for “Adjusting User Preference Settings”
Step 1:
With pump turned off, press and hold for 1.5 seconds the
+
and
the
ON/OFF
keys
at the
same time
. Pump will beep and then display the words BEEP HIGH or
BEEP LOW.
Press the
+
key to change to HIGH or the
-
key to change to LOW. Pump will give
a triple beep when the
+
or
-
keys are pressed and setting will change. If setting is
as desired, do not press either key.
Step 2:
Press the
PRIME
key. Display will read UNLK (unlocked) or LOCK (locked).
Press the
+
key to change to LOCK (locked) or the
-
key to change to UNLK
(unlocked). If setting is as desired, do not press either key.
Figure 2-23
MAX RATE Settings
Figure 2-24
MAX DOSE Settings
HIGH
LOW
LOCK
UNLK
MUTE
BEEP
LITE ON
LITE OFF
MAX RATE
MAX RATE
MAX DOSE
MAX DOSE