3. ALARMS, MESSAGES, AND INDICATIONS
© Copyright 2008 ZEVEX Incorporated
23
Why:
This alarm has occurred because pump has been in pause mode for 2
minutes.
To correct the problem:
Press the
RUN/PAUSE
key to silence alarm and
place pump in pause mode again. Program and use pump or press the
ON/OFF
key to turn it off.
SHUT DOOR (Figure 3-8)
Pump will sound a dual tone beep repeatedly and
display will read SHUT DOOR.
Why:
This alarm has occurred because the pump
door was opened while pump was running.
To correct the problem:
Press the
RUN/PAUSE
key to silence alarm and place pump in pause mode
then shut the door. Check door is latched securely.
If door will not stay closed, check door latch and
latch pocket on pump are not broken or cracked.
Messages
DOSE DONE (Figure 3-9)
This message indicates a feeding dose has been
delivered.
To clear pump:
Either press the
ON/OFF
key to
turn pump off, or press the
RUN/PAUSE
key to
put pump in pause mode before starting another
feeding.
Figure 3-8
SHUT DOOR Alarm
Figure 3-9
DOSE DONE Message