3.2
Requirements as to the Microscope Stand
VivaTome is permitted for operation in combination with the following microscope stands:
−
Axio Observer (all variants)
−
Axio Imager (all variants)
−
Axio Examiner (all variants, except for 2TV zoom tube)
The manufacturer does not assume liability for any use that is unrelated to the
intended use.
Any use of VivaTome in combination with microscope stands having an order number other
than that indicated in the present manual is not in accordance with the intended use.
Set up the microscope vibration-free on a vibration-damped table or on an anti-vibration plate.
The optimum performance of the complete system can be achieved only with a 100/0
deflection to the camera port with VivaTome mounted.
Long-term experiments require a stable ambient temperature to avoid any artifacts caused by
thermal expansion.
In addition to that, connection to the microscope requires a camera adapter 60N-C 1" 1.0x.