OPERATION
VivaTome
Controlling the Scanhead via AxioVision
Carl Zeiss
M60-2-0046 e 05/2012
423639-9000-701
21
4.3
Controlling the Scanhead via AxioVision
After switching on the scanhead, VivaTome can be controlled completely via the AxioVision software. In
this case, the device must be connected to the computer via the USB connector.
For operating the AxioVision software please see the
operating manual of AxioVision
.
4.4
Changing the Filter Module
Prior to changing the filter module, switch off the HXP 120 illuminator and disconnect the
VivaTome plug-in power unit from the mains socket outlet.
As the filter module and the camera adapter are properly matched to one another, they must
always be changed jointly. Filter sets and camera adapters which belong together have the same
ID number.
Change the filter module as follows:
•
Switch off the HXP 120 and disconnect the
plug-in power unit from the mains socket
outlet.
•
Disconnect the liquid light guide, power cable
and USB cable from VivaTome (see 3.7).
•
Remove VivaTome from the microscope and put
it on a soft surface.
•
If necessary, remove the supporting feet (see
chapter 3.5).
•
Screw the lifting screw (included in the filter set)
into the threaded hole (Fig. 4-2/
2
) of the base
plate (Fig. 4-2/
1
).
•
Lift the base plate by means of the lifting screw.
This requires some force due to the magnets
retaining the base plate.
Fig. 4-2
Removing the base plate