48
Operation
626140-9149-911
Operating manual TEMPAR console V9.0
3.2.22
"Configuration > Measuring Room >
Classification" menu
Fig. 42:
"Classification" menu
The "Classification" tab specifies the limit values for measuring room
monitoring (acc. to VDI/VDE guideline 2627). Here, it is only possible to
set the basic temperature for measuring room quality classes 2 to 4. In
measuring room quality class 1, the basic temperature is defined in the
guideline. All other values are available only for overview.
To use the special measuring room, it is first necessary to activate the
checkbox at the top left for "Use only special room calculation (Class S):".
This activates the special measuring room area. The values underlined in
blue can now be changed. While the special measuring room is active, a
check is only performed as to whether the measuring room is within the
specification for the special measuring room or not. Measuring room
classes 1 to 4 are not considered.