44
Operation
626140-9149-911
Operating manual TEMPAR console V9.0
– Identification number of the sensor in the sensor network
– Sensor description: Editable sensor description (e.g. Sensor top
door)
– DAkkS-calibrated (the information is saved in the sensor and
automatically transferred)
– Serial number of the gateway with which the corresponding sensor
unit has been paired (only when using TEMPAR wireless).
– The TEMPAR system is searched for connected components via
“Sensor network scan”
. If connected components are not listed,
increase the maximum network address.
– Clicking on
“Automatic sensor assignment”
causes the sensor
with the serial number to be assigned to the next free sensor name by
increasing number. All connected sensors can be easily assigned in this
way.
– The area
“Radio sensor pairing”
can only be seen if the TEMPAR
wireless gateway is present. Here, the gateway settings and pairing of
the sensor units with the gateway is undertaken. The following is
displayed for each connected gateway:
– Gateway serial number
– Start of pairing mode
– The type of measured value determination can be selected in the area
"
Temporal gradient configuration
".
Gliding window: The highest and lowest measured value in a time
window are used in the calculation.
Fixed interval measuring points: Only the current measured values and
the measured values recorded at the beginning of the interval are
used in the calculation.
– A time delay can be set in the area "
Time constants
" after which a
temperature change in the sensor is indicated. In this way a time
constant can be set in the sensors, that corresponds to a measuring or
testing instrument (e.g. coordinate measuring machine 25 minutes =
1500 seconds). If for example a time constant of 1500 seconds is set,
a temperature change of 1 K would only cause a change to the
measuring result of 1 K after 1500 seconds.