The "TEMPAR - Automatic Temperature Monitor" is intended solely for
the recording of measured values of interior spaces.
A feasible arrangement and number of sensor units must be selected. To
comply with standards, nine temperature sensors must be installed in a
measuring room so that they are at least 0.5 m away from the
walls/machine. One sensor is located in the centre of the space and the
remaining eight sensors are installed in the corners of the room at two
levels.
Technical specifications (See chapter "5 Technical data").
Proper usage also includes:
– Observing all notes in the manual.
– Maintenance of the calibration interval.
The following points must be complied when using TEMPAR wireless:
– Sensor units must not be exposed to heat or fire.
– Sensor units must not be exposed to mechanical impacts.
– If a battery leaks inside a sensor unit, do not allow the liquid to come
into contact with the skin or get into the eyes. If the battery fluid
touches the skin, wash the affected area with plenty of water and see
a doctor.
– No other chargers should be used other than those provided with the
devices specifically for their charging.
– Store sensor units in clean and dry conditions.