8 Operation
Instructions for Use
2660021169042 Rev. A 2018-03
8.5 Acquire Data (Scan)
86 / 246
2660021169042 Rev. A 2018-03
ð
The Acquire Screen opens.
1
2
3
4
5
6
7
2. Select the appropriate
scan (1)
under
OD
or
OS
.
ð
The chinrest moves into place for the selected scan.
3. Left-click on the center of the
pupil (2)
in the
viewport (3)
.
ð
The pupil target appears in the viewport.
4. If needed, move the
chinrest settings (4)
to position the
image in the
viewport (3)
.
5. Click
Auto Focus (5)
and ensure that the
pupil target (2)
remains centered.
NOTE
If the fundus image becomes dark after you click Auto Focus,
recenter the pupil target and adjust the image brightness.
6. To set the fixation target, click on a new location in the
fundus
image (6)
ð
The fixation target (green crosshair) moves to the location
selected.
7. Click
Optimize (7)
.
8. To adjust image brightness or make other fine adjustments for
any viewport, click the corresponding
Fine Adjustment (8)
tool and make adjustments. (B-Scan viewport indicated.)
Summary of Contents for PLEX Elite 9000
Page 1: ...PLEX Elite 9000 Version 1 7 Instructions for Use...
Page 54: ...Empty page for your notes...
Page 106: ...Empty page for your notes...
Page 174: ...Empty page for your notes...
Page 182: ...Empty page for your notes...
Page 210: ...Empty page for your notes...
Page 218: ...Empty page for your notes...
Page 224: ...Empty page for your notes...
Page 226: ...Empty page for your notes...
Page 228: ...Empty page for your notes...
Page 230: ...Empty page for your notes...
Page 245: ...Empty page for your notes...