![background image](http://html1.mh-extra.com/html/zeiss/plex-elite-9000/plex-elite-9000_instructions-for-use-manual_3362565204.webp)
13 Maintenance and Repair
Instructions for Use
2660021169042 Rev. A 2018-03
13.5 Power Table Fuse Replacement
204 / 246
2660021169042 Rev. A 2018-03
10. Replace the rear cover.
11. Your device may now be powered on.
13.5 Power Table Fuse Replacement
Two fuses are located just below the table power cord socket, near
the floor.
WARNING!
Failure to replace fuses with those of the same type and
rating
may create a fire risk.
u
Always replace fuses with the same type and rating.
CAUTION!
Not following these instructions to safely replace table fuses
could cause damage to the table or injury to personnel.
u
Always power down the table and unplug the power cord
before proceeding.
u
At all times, use the minimum force necessary to accomplish
each step so as to prevent damage or injury
1. Power down the system.
2. Unplug the table power cord at both ends, for safety and easy
access.
3. Grasp the housing while pressing down on the center tab of
the fuse holder and firmly pull out.
4. Check both fuse filaments for breakage.
Fuse type and rating for both system fuses: • FUSE, T 8.0A,
250V (100 – 120V ~) • FUSE, T 6.3A, 250V (220 – 240V ~)
5. Insert the new fuse(s) in the holder.
6. Slide the holder back into the housing until it snaps closed.
7. Plug in the power cord at both ends.
8. Your table is now ready to be powered on.
Summary of Contents for PLEX Elite 9000
Page 1: ...PLEX Elite 9000 Version 1 7 Instructions for Use...
Page 54: ...Empty page for your notes...
Page 106: ...Empty page for your notes...
Page 174: ...Empty page for your notes...
Page 182: ...Empty page for your notes...
Page 210: ...Empty page for your notes...
Page 218: ...Empty page for your notes...
Page 224: ...Empty page for your notes...
Page 226: ...Empty page for your notes...
Page 228: ...Empty page for your notes...
Page 230: ...Empty page for your notes...
Page 245: ...Empty page for your notes...